The RH Systems CGS-240 Humidity Generator is a system capable of continuous, high-accuracy humidity generation over a wide range of humidity, temperature, and flow rates.
The CGS-240 design is an RHS hybrid two-pressure two-temperature design based on fundamental NIST developed principles
Factory Tour
Specifications – CGS-240 | |
---|---|
Generating Range | |
Relative Humidity | 10 … 95 %RH (5 %RH optional) |
Dew/Frost Point | -25 … 70 °C DP, -23 … 0 °C FP |
Chamber Temperature | -0 … 72 °C (-10 & +85 °C extended ranges optional) |
Chamber Pressure | Ambient |
Flow Rate | 5 … 20 SLPM (5 … 50 SLPM optional) |
Accuracy (0 to 70 °C, 10 to 40 l/min) | |
Relative Humidity | The larger of ±0.1 %RH or ±0.5% of indicated RH value |
Dew/Frost Point | ±0.10 °C |
Temperature (TS, TC) | ±0.030 °C |
Pressure | ±0.02% of full scale |
Chamber Temp Uniformity | <0.035 °C |
Flow Rate | ±1 SLPM |
Integrated Sensors | |
Low Range Pressure (PL) | Ambient … 25 psia, ±0.005 psia |
High Range Pressure (PH) | Ambient … 150 psia, ±0.030 psia |
Saturator Temperature (TS) | -15 … 72 °C, ±0.03 °C |
Chamber Temperature (TC) | 0 … 72 °C, ±0.03 °C |
Chamber Liquid Temp (TCL) | 0 … 72 °C, ±0.10 °C |
Presaturator Temp (TP) | 0 … 75 °C, ±0.10 °C |
Flowmeter | 0 … 50 SLPM, ±1 SLPM |
Chamber | |
Dimensions | 15”H x 12”W x 12”D (380 x 300 x 300 mm) |
Type | Circulating air-jacket encased within liquid jacket |
Adjustable Shelf | Three shelf positions, one shelf supplied |
Access Ports | Three 50 mm diameter thermally controlled access ports |
Window Size | 14”H x 12”W x 12”D (370 x 300 x 300 mm) |
Window Thermal Control | Liquid jacketed, multi-pane, thermally controlled |
Chamber Light | Integrated light with on/off touchscreen control |
Integrated Control Interface | |
Operation Software | RHS Control Software |
Calibration Software | RHS Plateau Software |
Display | 18.5” Multi Touch Monitor with pinch/zoom/drag |
Gas Inlet Pressure | |
Using External Pressure Source | 120 psig max |
Using Supplied Air Compressor | 60 … 90 psig @ 20 l/min typical |
Environmental Conditions | |
Operating Environment | 15 … 30 °C, 20 … 60 %RH |
Storage Environment | 0 … 50 °C, <95 %RH non-condensing |
Power Requirements | |
Voltage | 208 … 240 VAC, 50/60 Hz, single phase |
Current | 20 Amps |